ClearanceBrief
Recall Z-0337-2016 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

NxStage Medical, Inc.

NxStage System One S Cycler -High Permeability Hemodialysis System Model no. NX1000-4

DateOct 29, 2015
StatusTerminated
ReasonUltrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed
Root causeSoftware design
ActionNxStage issued a Field Safety Notice (FSN), dated October 29, 2015, to clinic and home patients with a NX1000-3 or NX1000-4 cycler. For patients performing short daily hemodialysis, the FSN included instructions for steps to be taken so that the cycler and treatments are not affected by the software error. Patients performing nocturnal hemodialysis were instructed to contact their Center for a replacement NX1000-1 Cycler. A software update will be released to correct the error. Customers should complete and return the reply form. Contact NxStage Customer Service at 1-866-NXSTAGE (1-866-69
Quantity7 units
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140526NXSTAGE SYSTEM ONE
Nxstage Medical, Inc., May 19, 2014