Recall Z-0393-2016 ยท LZG Pump, Infusion, Insulin

Insulet Corporation

OmniPod, Insulin Management System (US) Catalog Number: PODZXP420 Product Usage: The OmnniPod Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger.

DateDec 15, 2015 (initiated Nov 2, 2015)
StatusTerminated
ReasonPod's needle mechanism fails to deploy or there is a delay in the deployment of the needle mechanism.
Root causeProcess change control
ActionInsulet issued on 11/2/15 an URGENT: Field Safety Notification via Email notification and Federal Express. Letter describes the problem that certain lots of the pod's needle failed to deploy or there is a delay in the deployment of the needle. Customers not responding to the email or Federal Express will receive additional mailing and/or follow up phone calls. Call Customer Care at 1-855-407-3729 if you have any questions regarding this Field Safety Notification.
Quantity26,230.9 boxes
DistributionWorldwide Distribution - US Nationwide and countries of Switzerland, Germany, and Israel.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122953OMNIPOD INSULIN MANAGMENT SYSTEM
Insulet Corporation, Dec 7, 2012