Recall Z-0230-2016 ยท CBK Ventilator, Continuous, Facility Use

Event Medical LTD

Inspiration Ventilator System, i Series and LS 12.1 LCD. The Inspiration 7i / 5i Ventilator System is to be used by healthcare professionals in hospitals or healthcare facilities and intra-hospital transport.

DateNov 19, 2015 (initiated Oct 13, 2015)
StatusTerminated
ReasonA component on the systems power board can fail in a manner which can cause the ventilator to shut down without alarming.
Root causeComponent design/selection
ActioneVent Medical sent an Urgent Field Safety Notice dated October 12, 2015, to all affected customers on October 13, 2015, to inform them that eVent Medical is recalling Inspiration Ventilators i Series and LS 12.1 LCD because a component on the systems power board can fail in a manner which can cause the ventilator to shut down without alarming. The letter provides the customers with the problems identified and the actions to be taken. Customers with questions are instructed to call (949) 900-1917, x232.
Quantity6441 units total (251 units in US)
DistributionWorldwide Distribution - US (Nationwide) and Internationally to: Algeria Argentina Australia Austria Bahrain Bangladesh Barbados Belarus Belgium Bolivia Canada Chile China Colombia Czech Rep Ecuador Egypt England Estonia Finland Germany Ghana Gibraltar Greece Guatemala Holland Hong Kong Hungary Indi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130178INSPIRATION VENTILATOR SYSTEM 5I/7I
eVent Medical, Ltd., Oct 29, 2013