Recall Z-0296-2016 ยท KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)

Fresenius Medical Care Renal Therapies Group, LLC

Fresenius Modified Mixer Motor (for 132 Gallon Dissolution Units), P/N 160075. For Use in the Preparation of acid concentrate for hemodialysis.

DateOct 23, 2015
StatusTerminated
ReasonDissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.
Root causeDevice Design
ActionFresenius issued an Urgent - Medical Device Correction notification, dated October 21, 2015, on 10/27/15. The letter identified the affected device and the reason for the recall. It also contained instructions (on how to perform the required wiring modification) along with a Reply Form. Customers are to return the form after completing the modification. Inquiries regarding the correction should be directed to Fresenius Technical Services at 800-227-2572, referencing the Field Action number FA-2015-01-W.
Quantity992 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131611FRESENIUS DRY ACID DISSOLUTION UNITS
Fresenius Medical Care North America, Aug 29, 2013