Recall Z-0273-2016 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips Electronics North America Corporation

ECG Out Cable,for use with a Philips ALS monitor/ defibrillator and bedside monitor for Synchronized cardioversion and Transcutaneous pacing. M1783A/M5526A 12-pin Sync cables, used with HeartStart MRx and HeartStart XL and HeartStart XL+

DateNov 13, 2015 (initiated Nov 19, 2014)
StatusTerminated
ReasonWhen a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTs) while connected to AC power: On the HeartStart MRx and HeartStart XL, EFT noise can be misinterpreted as an R-wave. On the HeartStart XL+, EFT noise can disable ECG monitoring, and potentially interrupt demand mode pacing.
Root causeDevice Design
ActionPhilips sent an Urgent - Field Safety Notice dated November, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips plans to take to remedy the issue. A replacement cable will be provided free of charge for all units affected by this issue. A Philips Healthcare representative will contact customers with affected devices to arrange for installation of the upgrade. Philips is asking customers to follow the: Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Saf
Quantity18,604
DistributionWorldwide Distribution - US (nationwide) and to the countries of : Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Ethiopia, Finland, France, Gabon, Germany, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Is
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031187HEARTSTART MRX, MODEL M3535A
Philips Medical Systems, Aug 27, 2003
K040404HEARTSTART MRX, MODEL M3535A
Philips Medical Systems, Inc., May 4, 2004
K051134HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A
Philips Medical Systems, Sep 8, 2005
K061707HEARTSTART MRX, INVASIVE PRESSURE & TEMPERATURE OPTIONS
Philips Medical Systems, Aug 11, 2006