Recall Z-0297-2016 ยท JQP Calculator/Data Processing Module, For Clinical Use

Roche Diagnostics Operations, Inc.

Bulk Loader Module for cobas p 512/612 Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis. In vitro diagnostic machine component.

DateNov 18, 2015 (initiated Sep 22, 2015)
StatusTerminated
ReasonPossible for sample tubes in the loader to be opened and contaminate the system as well as adjacent samples during processing.
Root causeDevice Design
ActionRoche Diagnostics Operations, Inc. sent an Urgent Medical Device Correction letter to affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were asked to retain the Urgent Medical Device Correction letter until a Roche Field Service Representative has installed the BLM modification kit. Consignees were also asked to complete the attached form and fax to 1-877-906-8982, ext 2015. Consignees with questions were instructed to call 1-800-428-2336. For questions regarding this recall call 317-521-3911.
Quantity1
DistributionDistributed to TX only.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.