Recall Z-0162-2016 ยท OSF Patient Specific Manual Orthopedic Stereotaxic System

Materialise N.V.

Signature Patient-Specific Surgical Guides; patient specific instruments used as a guide during surgical procedure for total and partial knee arthroplasty. Part of the Signature Personalized Patient Care System.

DateOct 23, 2015 (initiated Sep 23, 2015)
StatusTerminated
ReasonPotential inaccuracies in the alignment of the guides. Surgery was inaccurately planned because the patient image data was incorrectly processed.
Root causeUse error
ActionOn September 16, 2015 Materialise NV identified four cases where not all available image sets were used to design the case. On September 18, 2015, prior to the surgery dates Materialise informed (phone call) the Signature Services Manager of Zimmer Biomet that inaccuracies in the alignment of the knee might be observed with the use of the guides due to this issue. Biomet was asked to segregate the packages for these four cases and to return them to the Materialise US office. On September 23, 2015 this was followed up by e-mail notification letter with response forms. Biomet was requested to co
Quantity4
DistributionDistributed in MI, ID, TX, and in France.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102795SIGNATURE PLANNER; SIGNATURE GUIDES
Materialise NV, Feb 2, 2011