Recall Z-0345-2016 ยท FMT Warmer, Infant Radiant

DeRoyal Industries Inc

HYDRO-TEMP(R) Neonatal Temperature Skin Sensor, REF HNICU-32, 50 units per box, NON STERILE, Rx only.

DateNov 30, 2015 (initiated Oct 14, 2015)
StatusTerminated
ReasonDeRoyal manufactured four lots of HNICU-32, Hydro-Temp(R) Neonatal Temperature Skin Sensor incorrectly. The pins contained in the probe were soldered incorrectly. When probe is connected to the proper port, this defect results in the inability to read or display temperatures on the infant warming system.
Root causeProcess control
ActionDeRoyal issued their recall on 10/14/2015 to the end-user level by sending overnight letters to its direct accounts and hospitals. The consignees were instructed to destroy and discard the units. In the case of further distribution, the firm requested that the consignee either notify its customers of the recall or to submit a customer list for DeRoyal to contact them.
Quantity16 cases of 50 units (800 eaches)
DistributionDistributed in the states of VA, AL, FL, and CA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K925006HYDRO-TEMP NICU AND SKIN SURFACE TEMPERATURE SENSO
Deroyal Industries, Inc., Jan 31, 1994