Recall Z-0272-2016 ยท CAN Regulator, Pressure, Gas Cylinder

Precision Medical, Inc.

Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per minute.

DateNov 13, 2015 (initiated Oct 8, 2015)
StatusTerminated
ReasonThese lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for the 15 liter. For the 8 liter, instead of being in the OFF position, the regulator will be in the quarter liter setting (0.25).
Root causeError in labeling
ActionThe firm sent "Attention Field Corrective Action" letters, dated Oct 8, 2015, to customers. The letter identified the affected device and explained the reason for the recall. The letter provided instructions on how to check the regulator; and asked customers to call the number provided for a replacement if a regulator is labeled incorrectly.
Quantity315
DistributionDistributed in the US to the states of PA, MA, MO, MT, ID, MI, MN, ND, KY, NY, NC, CA, AL, AZ, LA, OH, TX, GA, IN, and OK.