Recall Z-0181-2016 ยท MUJ System, Planning, Radiation Therapy Treatment

Elekta, Inc.

Monaco - Radiation Treatment Planning used to make treatment plans for patients with prescriptions for external beam radiation therapy.

DateOct 30, 2015 (initiated Oct 16, 2015)
StatusTerminated
ReasonUnintended update of Dose and MU and Incorrect Assignment of Bolus.
Root causeOther
ActionElekta sent Important Field Safety Notice 382-01-MON-003 to all customers on 10/16/2015. The notice informs users of the specific product and version numbers affected by both issues. The affected devices for both issues are those running Monaco 5.10.01. Customers are instructed to complete and return acknowledgement form. Solution/fix has been released in Monaco patch 5.10.02 and all affected users will be made aware of its availability by the end of October 2015.
Quantity154
DistributionCA, IN, MI,MO, TX, WI, Australia, Canada, Germany,m Greece, India, Netherlands, New Zealand, Turkey and United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151233Monaco
Impac Medical Systems, Inc., Jul 7, 2015