Recall Z-0339-2016 ยท JGJ Photometric Method, Magnesium

Siemens Healthcare Diagnostics, Inc.

Dimension Vista System (MG) ; Flex reagent cartridge Catalog number K3057, SMN 10445158. For the quantitative measurement of magnesium in human serum, plasma and urine on the Dimension Vista System.

DateNov 25, 2015 (initiated Sep 30, 2015)
StatusTerminated
ReasonErroneous low results on a small number of reagent wells. There is the potential for under-recovery on certain wells for quality control (QC) and patient sample results. The under-recovery ranged from -0.3 mg/dL [0.12 mmol/L] to -1.6 mg/dL [0.66 mmol/L]; equivalent to a maximum negative bias of -88%.
Root causeNonconforming Material/Component
ActionAn Urgent Medical Device Recall dated September 2015 was provided to all Dimension Vista MG customers who received the affected lot to notify them of the issue, the potential risk to health, and actions to be taken by customers. 1. An Urgent Medical Device Recall was provided to all customers who received the affected lot to notify them of the following: Siemens has received customer complaints for Dimension Vista MG lot 15063BA, regarding sporadic depressed serum/plasma and QC results. Discontinue use and discard their remaining inventory of the affected lot. Review the letter with their Med
Quantity3601
DistributionDistributed US (nationwide) including DC and the states of IL, CA, GA, LA, TX, TN, FL, SC, IA, TN, DE, NJ, WA, MS, KY, NY, WI, OH, OH, MO, UT, KS, VA, AR, WV, MA, PA, AL, CO, IN, MT, NM, ME, AZ, NM, MO, and NC, and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061655DIMENSION VISTA FLEX REAGENT CARTRIDGES
Dade Behring, Inc., Jul 10, 2006