Recall Z-0798-2016 ยท DQO Catheter, Intravascular, Diagnostic

Cook Inc.

Shuttle Select Slip-Cath Catheter. Angiographic catheter. For use in angiographic procedures as a conduit for the delivery of contrast media and can help physicians diagnose occlusion or stenosis. The catheters are designed for percutaneous introduction into the vascular system over an appropriately sized wire guide.

DateFeb 24, 2016 (initiated Oct 7, 2015)
StatusTerminated
ReasonReports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result. Recall expanded 10/07/2015 to include additional catheters.
Root causeComponent design/selection
ActionOn 7/2/2015 Cook Incorporated sent, URGENT: MEDICAL DEVICE RECALL notifications dated July 2, 2015, to affected customers with instructions for returning the affected product. All customers were notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Should you have any questions or concerns, please do not hesitate to contact Cook Medical Customer Relations at (800) 457-4500 or (812) 339-2235. Firm issued press release 8/3/2015.On 10/09/201
Quantity9519
DistributionWorldwide Distribution - US (nationwide including Puerto Rico and US Virgin Islands), Algeria, Antigua, Armenia, Austria, Australia, Bahamas, Belgium, Bangladesh, Bulgaria, Bahrain, Brazil, Brunei Darussalam, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122937SLIP-CATH BEACON TIP CATHETER MODEL SCBR (PRODUCT PREFIX), SHUTTLE SELECT SLIP-CATH CATHETER MODEL SCBR (PRODUCT PREFIX)
Cook, Inc., Dec 14, 2012