Recall Z-0280-2016 ยท LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy

Cyberonics, Inc

Cyberonics Vagus Nerve Stimulator (VNS) Therapy AspireSR Generator, Model 106

DateNov 17, 2015 (initiated Oct 7, 2015)
StatusTerminated
ReasonCertain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit potential for decreased battery longevity.
Root causeManufacturing material removal
ActionCyberonics sent written recall notifications beginning October 14, 2015, to all affected customers. The letter identified the product the problem and the action to be taken by the customer. The notice also directs customers to contact Cyberonics to schedule additional on-site troubleshooting. The on-site troubleshooting proposed by Cyberonics will utilize a feature included within the Model 250 Version 9.0 Programming Software as an alternate method of verifying heart rate and determining that the appropriate heart beat sensitivity setting is programmed. The Version 9.0 software displays a
Quantity27 units
DistributionUS Distribution to the states of : OR, NY, GA, PA, CA, NC, KS, FL, KY, MI, TX, IL, OH, IN, UT, CO and WV.