Recall Z-0182-2016 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Siemens Medical Solutions USA, Inc.

ACUSON SC2000 ultrasound systems model 10433816 Imaging system that provides the ability to measure anatomical structures and calculation packages that provide info for clinical diagnosis purposes.

DateOct 30, 2015 (initiated Oct 7, 2015)
StatusTerminated
ReasonSiemens will provide all customers a new version of system software (VA35E or VB10C depending on the hardware configuration). Hardware fixes will be implemented to address situations where the metal component used to engage the wheel lock becomes disconnected, and where the system locks up due to misalignment between the bezel and MPI board, in all units that have not already been corrected.
Root causeDevice Design
ActionOn August 19, 2015, Siemens Ultrasound began sending a customer notification to users, advising them to correct the uppercase/lowercase differences in the patient name at the time of registration until the issue is resolved with a future software release. Siemens believed that the other issues in this correction will begin approximately November 30, 2015 and December 31, 2015. Field service reps will visit sites to install software updates or hardware corrections.
Quantity2,099 units
DistributionWorldwide Distribution -- US and Albania, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia Herzeg., Brazil, Brunei, Bulgaria, Canada, Colombia, Croatia, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt, El Salvador, Finland, Fr
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123622ACUSON
Siemens Medical Solutions USA, Inc., Dec 5, 2012