Recall Z-0438-2016 ยท OYC Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor

Medtronic MiniMed Inc.

MiniMed 640G, 3.0mL mmol/L, 3.0mL mg/dL, Model No. MMT-1711 and MMT-1712 The MiniMed 620G and MiniMed 640G insulin pump systems are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. In addition, the pump system is indicated for continuous monitoring of glucose levels in the fluid under the skin, and p

DateDec 17, 2015 (initiated Sep 25, 2015)
StatusTerminated
ReasonThe pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being delivered. Possible cause is either lack of greasing or deficient greasing in a ball bearing assembly of the motor.
Root causeNonconforming Material/Component
ActionMedtronic sent an Urgent Medical Device Recall Notification dated September 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letters stated that the pump drive motors may experience a malfunction which would result in a pump error message alarm notifying the patient that insulin is no longer being delivered. The letters also inform the customers of what actions are to be taken. Customers with questions are instructed to contact Medtronic MiniMed's customer support team. For further questions please call (81
Quantity1052 units
DistributionInternationally Distribution only to the countries of : Australia, Austria, Belgium, Chile, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, India, Ireland, Italy, Japan, Kenya, Luxembourg, Namibia, Netherlands, Norway, Slovenia, South Africa, Spain, Sweden, Switzerland, United K
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.