Recall Z-0215-2016 ยท CGA Glucose Oxidase, Glucose

ELITech Clinical Systems SAS

Eon 300 system includes the following product reference numbers and names: Reference no. 79345 Glucose (GPSL) Reference no. 79450 Uric acid (AUVD) Reference no. 79435 Triglycerides (TGML) Reference no. 79295 Cholesterol (CHVD) Reference no. 79301 Cholesterol HDL (HDLL) Reference no. 79310 Cholesterol, LDL (LDLL)

DateNov 5, 2015 (initiated Sep 22, 2015)
StatusTerminated
ReasonELITech clinical Systems reagents based on Trinder reaction used on the Eon 300 instrument system are recalled due to potential drug interference. No complaints were received.
Root causeUnder Investigation by firm
ActionThe firm sent recall letters or Technical Bulletins to US consignees on 9/22/15 via registered mail. Technical Bulletin #1022 (for EON 100 users) Technical Bulletin #2008 (for EON 300 users) Technical Bulletin #5064 (for ENVOY users) Technical Bulletin #3009 (for Selectra users) ELITechGroup is not requesting return of the affected product. The method sheets and package inserts will be updated with the new interference claims. When available, the revised labeling will be provided to the customer for customer's use until the new IFU is included in the product packaging. Consignees are reques
Quantity612 total
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100525ELITECH CLINICAL SYSTEMS GLUCOSE PAP SL ELITECH CLINICAL SYSTEMS ELICAL 2, ELITECH CLINICAL SYSTEMS ELITROL 1 AND 2
Seppim S.A.S., Dec 15, 2010