Recall Z-0203-2016 ยท KWQ Appliance, Fixation, Spinal Intervertebral Body

Biomet Spine, LLC

MaxAn 4.0mm and 4.5mm fixed screws, used in the MaxAn Anterior Cervical Plate System. The MaxAn Anterior Cervical Plate System is a decompression-based technique for cervical spine stabilization.

DateNov 3, 2015 (initiated Sep 29, 2015)
StatusTerminated
ReasonThe packaging of 4.5 mm diameter MaxAn screws identifies/labels the screws as 4.0 mm diameter screws.
Root causeLabeling mix-ups
ActionZimmer Biomet Spine sent an Urgent Medical Device Recall Notice dated September 29, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. If the device has already been implanted, written confirmation of this will be required and the necessary parties will be contacted. They will store returned product in quarantine in an isolated location until they can undergo inspection to verify if they are conforming or not and then will be destroyed or retained. Necessary parts will be destroyed so that it is visually obvious t
Quantity17 units
DistributionUS Distribution to the states of : NY, MO, and FL
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080646C-TEK MAXAN ANTERIOR CERVICAL PLATE SYSTEM
Biomet Spine, May 2, 2008
K133518MAXAN ANTERIOR CERVICAL PLATE SYSTEM
Biomet Spine (Aka Ebi, LLC), Nov 5, 2014