Recall Z-0255-2016 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare

GE Healthcare, SIGNA 3.0T. Product Usage: Magnetic resonance system for diagnostic imaging. The indications for use for the 3.0T Signa VH/i (Signa 3.0T MR System) Transmit/Receive Body Imaging Coil expands the imaging capability of the 3.0T Signa VH/i MR Imaging System. The Transmit/Receive Body Imaging Coil is intended for imaging of the Neck, Spine, Abdomen/Thorax and the extremities.

DateNov 10, 2015 (initiated Aug 31, 2015)
StatusTerminated
ReasonWhen performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.
Root causeSoftware design
ActionGE Healthcare sent an Urgent Medical Device Correction letter dated August 31, 2015. The letter was addressed to Hospital Administrators/Risk Managers, Radiology Department Managers, & Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity19
DistributionWorldwide Distribution - US: Nationwide including DC, PR and GU and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOSNIA HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMAN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013636SIGNA INFINITY MR SYSTEM WITH EXCITE TECHNOLOGY
GE Medical Systems, Jan 24, 2002