Recall Z-0112-2016 ยท MUJ System, Planning, Radiation Therapy Treatment

Elekta, Inc.

Monaco Radiation Treatment Planning System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy.

DateOct 7, 2015 (initiated Sep 15, 2015)
StatusTerminated
ReasonDose and MU are incorrect when CT images are viewed from the head, and, when using multiple prescriptions with forced densities.
Root causeDevice Design
ActionThe firm issued an URGENT IMPORTANT FIELD SAFETY NOTICE on 09/15/2015. The letter identified the affected device, as well as the reason for the recall, and clinical impact. Customers were informed that both issues would be resolved in Monaco Patch Release 5.10.01. The attached acknowledgement form should be completed, signed, and returned to a local Elekta Office or representative within 30 days. Questions regarding the notice should be directed to a local Elekta office.
Quantity154 units
DistributionWorldwide Distribution -- US, including the states of CA, IN, MI, MO, TX and WI; and, the countries of Australia, Canada, Germany, Greece, India, Netherlands, New Zealand, Turkey, and the UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151233Monaco
Impac Medical Systems, Inc., Jul 7, 2015