Recall Z-0021-2016 ยท JAK System, X-Ray, Tomography, Computed
Siemens Medical Solutions USA, Inc
SOMATOM Definition AS, intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.
| Date | Oct 5, 2015 (initiated Aug 14, 2015) |
| Status | Terminated |
| Reason | software bug issues for SW-Version VA48A_SP0. The following safety issues were resolved: 1) Correction to improve visual warning and error indication son the gantry display. 2) Correction to improve acquisition data in order to optimize image quality. 3) Correction to improve robustness and general system behavior in some exception handling procedures. 4) Correction to improve auto post processin |
| Root cause | Software design |
| Action | A customer advisory notice, dated 8/14/15, was sent to direct accounts advising them of the issue to observe the safety notice and comply with the corresponding measure until the update has been fully completed. |
| Quantity | 113 total |
| Distribution | Nationwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K143400 | SOMATOM Definition AS/AS+ Siemens Medical Solutions USA, Inc., Apr 8, 2015 |