Recall Z-2823-2015 ยท OLO Orthopedic Stereotaxic Instrument

Medtronic Navigation, Inc.

P/N: 9735023, DRIVER 9735023 SOLERA 5.5/6.0 MAS; P/N: 9735024, DRIVER 9735024 SOLERA 5.5/6.0 MAS CAN; P/N: 9735025, DRIVER 9735025 SOLERA 5.5/6.0 RMAS; P/N: 9735026, DRIVER 9735026 SOLERA 5.5/6.0 RMAS CAN; P/N: 9735027,DRIVER 9735027 SOLERA 5.5/6.0 FAS/SAS. The Medtronic Navigated Solera Screwdrivers are reusable devices that are designed for use in spine surgery to assist in the placement of scr

DateSep 30, 2015 (initiated Sep 14, 2015)
StatusTerminated
ReasonSpecific use conditions can increase the possibility of a screwdriver tip deforming or breaking during surgery.
Root causeLabeling design
ActionMedtronic sent an Urgent Medical Device Correction Notification letter dated September 14, 2015, to all affected consignees via FedEx. Revised Instructions for Use were provided to add additional warnings related to the careful inspection of the instruments and the importance of knowledge of operating procedures, patient selection, and product information. Consignees with questions were instructed to to call the local Medtronic Navigation Technical Services Representative or call 1-800-995-9709.
Quantity2,346
DistributionWorldwide Distribution - US (nationwide) and Internationally to Canada; United Arab Emirates; Australia; Belgium; Switzerland; China; Germany; Denmark; Spain; Finland; France; Great Britain; Greece; India; Israel; Italy; Japan; Netherlands; New Zealand; Poland; Portugal; Sweden; Singapore; Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K124004NAVIGATED CD HORIZON SOLERA SCREWDRIVERS, TAPS, ILIAC TAPS, LEGACY TAPS
Medtronic Navigation, Inc., Mar 22, 2013