Recall Z-0263-2016 ยท IZI System, X-Ray, Angiographic

NeuroLogica Corporation

Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.

DateNov 10, 2015 (initiated Sep 14, 2015)
StatusTerminated
ReasonWhen the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.
Root causeDevice Design
ActionAn Urgent - Field Safety Notice identifying the affected products, problem involved, root cause, and potential hazards, was sent to customers. In the event the issue occurs, users can release the drive handle or press the emergency switch and the unit will stop. Then, reboot the system. If unintended movements occur, please contact local representative. To correct the issue, service engineers will visit to apply the corrective action. For further information or support concerning this problem, contact local representative.
Quantity7 devices
DistributionDistributed to the states of NC, OH, MA, and MO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142492GM60A-32S, GM60A-40S
Samsung Electronics Co., Ltd., Dec 17, 2014
K150097GM60A-32S & GM60A-40S
Samsung Electronics Co., Ltd., Feb 19, 2015