Recall Z-2811-2015 ยท GXN Plate, Cranioplasty, Preformed, Non-Alterable

Synthes (USA) Products LLC

Synthes Cranial Flap Tube Clamp and Crimping Device Product Usage: The Synthes Cranial Flap Tube Clamp System is intended to reattach a cranial bone flap to the surrounding cranium after a craniotomy procedure.

DateSep 23, 2015 (initiated Aug 31, 2015)
StatusTerminated
ReasonThe Synthes Cranial Flap Tube Clamp and Crimping Device for Cranial Tube Clamp had been labeled MR Safe although they do not have the testing protocols currently required to designate them as MR Safe.
Root causeLabeling False and Misleading
ActionDePuy Synthes sent an URGENT:FIELD SAFETY NOTIFICATION / MEDICAL DEVICE LABELING CORRECTION letter, dated August 31, 2015 to affected customers. The letter identified affected product, problem and actions to be taken. Customers were instructed to: "Update and review the package insert (GP1345-D). " Discard outdated package insert noted in the table above. " Forward this Field Safety Notification to anyone in your facility that needs to be informed, especially personnel that conduct MR testing. " If the package insert has been forwarded to another facility, contact that facility. " Complete
Quantity105882
DistributionUS Nationwide and internationally, including Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K992000SYNTHES CRANIAL FLAP TUBE CLAMP
Synthes (Usa), Jul 20, 1999