Recall Z-2777-2015 ยท LPB Cardiac Ablation Percutaneous Catheter

Medtronic Inc. Cardiac Rhythm Disease Management

Medtronic CryoConsole, models 106A3, 106E2, and 106A2-K For use in performing cardiac ablation procedures.

DateSep 17, 2015 (initiated Sep 1, 2015)
StatusTerminated
ReasonAn issue with a USB memory component contained within a subset of CryoConsoles can result in extended procedure time.
Root causeComponent design/selection
ActionMedtronic began hand delivering an Urgent Medical Device Correction letter on September 1, 2015, to all affected customers. The letter described the issue. Customers were advised to continue using the device, per the operator's manual, and if a system notice appears, they are to carefully read and follow the instructions in the screen of the console. Medtronic personnel will track contact with the customer via Customer Contact Record. Customers with qeustions were instructed to contact their Medtronic AF Solutions representative. For questions regarding this recall call 1-877-464-2796 Tec
Quantity103 units
DistributionWorldwide Distribution - US (nationwide) and Internationally to Algeria, Australia, France, Germany, Greece, Iceland, Ireland, Israel, Italy, Lebanon, Malaysia, Martinique, Netherlands, Poland, Romania, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.