ClearanceBrief
Recall Z-2813-2015 ยท KHP Acid, Lactic, Enzymatic Method

Siemens Healthcare Diagnostics Inc

Siemens RAPIDPoint 500 Measurement Cartridge Lactate (750 tests) Part Number: 10491449 Product Usage: The RAPIDPoint 500 system is designed for professional use in a point-of-care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are test

DateSep 24, 2015 (initiated Aug 24, 2015)
StatusTerminated
ReasonRAPIDPoint 500 measurement cartridge serial numbers 2517102517 through 2519000012, negative sodium bias on both QC and patient results
Root causeProcess control
ActionSiemens Healthcare sent an Urgent Field Safety Notice letter dated August 2015 to all affected customers. The notice identified the affected product, problem and actions to be taken. Customers were instructed to complete and return the Field Correction Effectiveness Check Form attached. For questions contact your Siemens Customer Care Center or ther local Siemens technical support representative.
Quantity2079 cartridges
DistributionWorldwide Distribution - US Nationwide and the countries of Canada, Albania, Austria, Australia, Bosnia Herzeg., Bangladesh, Belgium, Brazil, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Denmark, Estonia, Spain, Finland, France, Great Britain, Georgia, Greece, Croatia, Hun
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113216RAPIDPOINT 500 BLOOD GAS ANALYER
Siemens Healthcare Diagnostics, May 3, 2012