Recall Z-2755-2015 ยท NAY System, Surgical, Computer Controlled Instrument
Intuitive Surgical, Inc.
Crocodile Grasper; used with the da Vinci Si Surgical System (IS3000); Model numbers: 428059-12, 428059-13. The Intuitive Surgical da Vinci Single-Site Instruments and Accessories used with the da Vinci Si Surgical System (IS3000) are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, appr
| Date | Sep 10, 2015 (initiated Aug 8, 2015) |
| Status | Terminated |
| Reason | In certain circumstances involving select da Vinci Si Single-Site instruments, the jaws of grip actuated instruments may become fixed in a closed position. |
| Root cause | Device Design |
| Action | Intuitive Surgical sent an Urgent Device Correction letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to do the following: 1. Ensure that all affected personnel are fully informed of this notice. Forward this notice to your Risk Manager, OR Director, Purchasing Manager, Biomedical Engineering staff and members of your medical staff who perform da Vinci Surgery procedures. 2. Replace any instrument that shows signs of damage and contact Intuitive Surgical Customer Service. 3. Complete and ret |
| Quantity | 15,834 total units |
| Distribution | Worldwide Distribution - US (nationwide) and Internationally to Australia, Belgium, Brazil, Canada, Chile, China, Cyprus, Denmark, France, Germany, Greece, Israel, Italy, Monaco, Netherlands, Panama, Romania, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey and Unite |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K120215 | SINGLE-SITE MEDIUM-LARGE CLIP APPLIER, SINGLE-SITE CADIERE GRASPER, SINGLE-SITE FUNDUS GRASPER, SINGLE-SITE CROCODILE Intuitive Surgical, Inc., Apr 30, 2012 |