Recall Z-2734-2015 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

da Vinci Xi Surgical System P4 software;with Xi Stapler; model number IS4000 : General and Plastic surgery: The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick

DateSep 3, 2015 (initiated Aug 17, 2015)
StatusTerminated
ReasonComplaints regarding the Exposed Knife Blade recoverable fault for the Xi Stapler with Xi System P4 version software.
Root causeSoftware design
ActionIntuitive sent an Urgent Medical Device Correction letters dated August 21, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. For questions contact Clinical Sales or ISI Customer service at 800-876-1310 Option 3.
Quantity198 affected devices
DistributionWorldwide Distribution - US Nationwide and in the countries of Italy and Norway.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131861DA VINCI SURGICAL SYSTEM, ENDOWRIST INSTRUMENTS AND ACCESSORIES
Intuitive Surgical, Inc., Mar 28, 2014