Recall Z-0149-2016 ยท LBI Unit, Neonatal Phototherapy

Natus Medical Incorporated

Natus neoBLUE 3 Phototherapy system or Neonatal Phototherapy, Catalogue Part Number: 010066; Top Assembly Part Number: 001103

DateOct 21, 2015 (initiated Aug 5, 2015)
StatusTerminated
ReasonThe neoBLUE Instruction For Use (IFU) and Service Manuals for the neoBLUE 3 were revised to show the cleared intensity statement settings so the output intensity aligns precisely with the information listed in the cleared 510(k) submission.
Root causeNo Marketing Application
ActionNatus Medical starting sending the Product Labeling Change NeoBlue LED Phototherapy System letter, dated August 2015, to all affected customers. The firm also attached the electronic copy of the revised User Manual with the letter. Customers are informed that the user and service manuals related to intensity for the neoBLUE 3 system are revised. Also, while the neoBLUE system does allow the light output to be adjusted with the potentiometers to accommodate different treatment heights, the firm removed the phrase "adjustable to >50 W/cm/nm" from the specifications. If you would like a CD dis
Quantity4560 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022196NATUS BLUE LIGHT PHOTOTHERAPY UNIT
Natus Medical, Inc., Sep 19, 2002