ClearanceBrief
Recall Z-0153-2016 ยท KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Innovative Sterlization Technologies LLC

OneTray Sealed Sterilization Containers Product Usage: A rigid sterilization container used to steam sterilize medical devices

DateOct 22, 2015 (initiated Feb 3, 2014)
StatusTerminated
ReasonThe wrong sterilization time was on the label. The label lists Steam Gravity 10 - 34 minutes at 270 degrees F instead of Steam Gravity 17 - 34 minutes at 270 degrees F.
Root causeProcess control
ActionThe firm sent an Urgent Medical Device Recall letter dated September 14, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to fill out the attached response SASE mailer and return to or fax to the address/number identified below: Innovative Sterilization Technologies, LLC, 7625 Paragon Road, Suite A Dayton, OH 45459, 937-286-3571 (C), 937-630-4346 (F), 937-619-0138 x 202 (Office), dbillman@ISTSterilization.com
QuantityM2104 - 25 units; M2408 - 84 units
DistributionUS Nationwide Distribution in the states of IL, IN, LA, OH, PA, TX & WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052567ONE TRAY SEALED STERILIZATION CONTAINER
Sterile Containment Technology, LLC, Mar 13, 2006