ClearanceBrief
Recall Z-2826-2015 ยท FPO Stretcher, Wheeled

Ferno-Washington Inc

Ferno iNX Wheeled Stretcher, Integrated Patient Transport & Loading System Product Usage: Stretcher for pre-hospital patient transport.

DateOct 1, 2015 (initiated Aug 3, 2015)
StatusTerminated
ReasonHole elongation in the telescoping load frame channel. A hole elongation could affect the ability of the retention pin to stay seated which is necessary to preclude movement of the telescoping load frame.
Root causeDevice Design
ActionFerno began calling their customers on 8/3/2015. The firm informed their customers of the following: To provide optimal Customer Experience and minimal disruption to your service, the iNX field action involves the following steps: " Exchange of your current iNX with a new iNX; " Participation in a brief training session; and " Inspection of your Ferno Model 175 antler and rail fastening system as requested by the Dept. of Transportation. As you await delivery of your new iNX, it is important you ensure proper function of the fastening system and iNX telescoping load frame.
Quantity72 units
DistributionWorldwide Distribution - US Nationwide in the states of: CO, IA, IL, IN, KY, MD, MN, MO, NC, NJ, NY, OH, OK, OR, TX, VA, WA & WY and the countries of CANADA, AUSTRALIA, UNITED KINGDOM & NEW ZEALAND.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.