Recall Z-2793-2015 ยท LHC Warmer, Irrigation Solution

Ecolab Inc

Microtek Medical, Inc. ORS Fluid Warming and Slush Drapes, Model #'s ORS-100, ORS-110, ORS-130, ORS-300, ORS-301, ORS-320, ORS-321, ORS-325, ORS-330, ORS-331 and ORS-2200. The ORS Fluid Warming and Slush Drapes are individually wrapped and placed in poly-Tyvek pouches with a label insert, then packaged as 24 per case in a labeled corrugated shipper box. Intended for use during various surgerie

DateSep 18, 2015 (initiated Aug 18, 2015)
StatusTerminated
ReasonThe sterile packaging may contain small (less than 250 microns wide) channels in the seal of the pouch. If channels are present, the package integrity is compromised and there is a remote chance that a non-sterile drape may be used during a surgical procedure.
Root causePackage design/selection
ActionConsignees were sent on 8/18/2015 an Ecolab "Urgent Medical Device Recall" letter dated August 18, 2015. The letter was addressed to Recall Coordinator or Risk Manager. The letter described the problem and the product involved in the recall. Advised consignees to quarantine all affected lots, and notify their customers of the recall which is to be carried out to users level. Requested consignees to complete and return the enclosed Recall Response Form. For questions they can contact Customer Service at 1-800-824-3027 or e-mail customerservice@microtekmed.com.
Quantity5,172,816 units
DistributionWorldwide Distribution -- US, including DC and PR; and, the countries of SAUDI ARABIA, LEBANON, GERMANY, HONG KONG, SINGAPORE, CANADA, and JAPAN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021288ORS-1000LD
O.R. Solutions, Inc., Jun 27, 2002
K023282ORS-3000LD
O.R. Solutions, Inc., Nov 13, 2002
K142080NON-STERILE ORS WARNING DRAPES/SLUSH+WARMER DISC-DRAPE/SLUSH+WARMER PALTE-DRAPE /FLUID WARMER DRAPE,SKIRTED
Ecolab, Inc., Aug 28, 2014