Recall Z-0408-2016 ยท JAK System, X-Ray, Tomography, Computed
Philips Electronics North America Corporation
Philips Healthcare Ingenuity CT Computed Tomography X-ray system
| Date | Apr 14, 2015 |
| Status | Terminated |
| Reason | During scans with specific protocol steps, the software unexpectedly sets the Axial ( 2 axis ) scan length to a different length than the user set in the plan. |
| Root cause | Radiation Control for Health and Safety Act |
| Action | Philips Healthcare Planned Action: 1. Philips Healthcare will contact customers and initiate a software update to correct the defect. 2. Philips Healthcare engineers will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that Philips Healthcare will without charge, remedy the defect or bring the product into compliance. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter i |
| Quantity | 105 |
| Distribution | Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Gibraltar, Great Britain, Greece, Hong Kong, India, Indone |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K033326 | PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM Philips Medical Systems, Inc., Oct 29, 2003 |