ClearanceBrief
Recall Z-0405-2016 ยท JAK System, X-Ray, Tomography, Computed

Philips Electronics North America Corporation

Philips Healthcare Brilliance 64 Computed Tomography X-ray system

DateApr 14, 2015
StatusTerminated
ReasonDuring scans with specific protocol steps, the software unexpectedly sets the Axial ( 2 axis ) scan length to a different length than the user set in the plan.
Root causeRadiation Control for Health and Safety Act
ActionPhilips Healthcare Planned Action: 1. Philips Healthcare will contact customers and initiate a software update to correct the defect. 2. Philips Healthcare engineers will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that Philips Healthcare will without charge, remedy the defect or bring the product into compliance. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter i
Quantity7
DistributionWorldwide Distribution - USA (nationwide) Distribution and to the countries of : Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Gibraltar, Great Britain, Greece, Hong Kong, India, Indone
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033326PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM
Philips Medical Systems, Inc., Oct 29, 2003