ClearanceBrief
Recall Z-0015-2016 ยท JAK System, X-Ray, Tomography, Computed

Toshiba American Medical Systems Inc

Toshiba Aquilion CT System TSX-301B

DateOct 9, 2015 (initiated May 8, 2015)
StatusTerminated
ReasonIf two specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved.
Root causeRadiation Control for Health and Safety Act
ActionToshiba America Medical Planned Action to bring defect into Compliance: a. A notification will be sent to all consignees with affected systems in compliance with 21 CFR1 003.21 upon approval by the Secretary. b. To correct this issue revised software will be installed on each system to prevent this occurrence from happening. Each consignee will be contacted by their Toshiba service representative for an appointment when the revised software becomes available. c. Each consignee will be advised to do the following until the corrective measures have been taken, i.e., revised software installed
Quantity459
DistributionUS Distribution including Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120546AQUILION PREMIUM
Toshiba Medical Systems Corporation, Japan, Jun 13, 2012