Recall Z-2420-2015 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

Medtronic Perfusion Systems

Medtronic EOPA Arterial Cannula, Model # 77620. Lot: 2015060544 intended to deliver oxygenated blood to the heart during cardiopulmonary bypass procedures for up to 6 hours in duration.

DateAug 19, 2015 (initiated Jul 24, 2015)
StatusTerminated
ReasonEOPA Arterial Cannula Devices in this lot were shipped without the guidewire.
Root causePackaging process control
ActionOn July 24, 2015 Medtronic sales representatives began verbally notifying the customers of the issue, requesting quarantine and return of un-used, affected product Consignees were sent a Medtronic "Urgent Medical device Recall" letter dated August 2015. The letter was addressed to Risk Manger and Health Care Professional. The letter described the problem and the product involved in the recall. Advised consignees to quarantine and remove the product from their inventory and return it to Medtronic. Requested consignees to complete and return the Customer Confirmation Certificate via email to
Quantity221 (111 US, 110 OUS)
DistributionWorldwide distribution. US nationwide including: FL, GA, IL, MI, NY, NC, PA, WV; Germany, Hong Kong, and Italy.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031037EOPA ELONGATED ONE-PIECE ARTERIAL CANNULA AND GUIDEWIRE
Medtronic Perfusion Systems, May 13, 2003