Recall Z-2718-2015 ยท PDU Catheter For Crossing Total Occlusions

Volcano Corporation

Volcano Pioneer Plus Re-Entry Catheter; Product Code PPLUS20, Catheter for Crossing Total Occlusions. Cardiovascular: The Volcano Pioneer Plus Intravascular Ultrasound Re-Entry Catheter (Pioneer Plus) is intended to facilitate the placement and positioning of catheters within the peripheral vasculature.

DateSep 2, 2015 (initiated Jul 22, 2015)
StatusTerminated
ReasonDue to a manufacturing defect, there is the possibility that a small wire could extend through the catheter shaft of the Pioneer Plus intravascular Ultrasound Guided Re-Entry Catheters.
Root causeUnder Investigation by firm
ActionThe Volcano Corporation sent an Urgent Medical Device Recall letter dated July 22, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately quarantine and return all affected product in their inventory. Customers were also asked to complete the attached form and fax back to 866-723-2482. Customers with questions were instructed to call 866-861-8923. For questions regarding this recall call 978-439-3586.
Quantity5042 devices
DistributionWorldwide Distribution - US (nationwide) Canada, Europe, Saudi Arabia, Iran, Egypt, and Qatar.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081804PIONEER PLUS CATHETER
Medtronic, Inc., Aug 5, 2008
K101777PIONEER PLUS CATHETER, MODEL PLUS 120
Medtronic Vascular, Jul 23, 2010