Recall Z-2733-2015 ยท MSW System, Test, Thyroglobulin

Beckman Coulter Inc.

Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobulin levels in human serum and plasma using the Access Immunoassay Systems.

DateSep 3, 2015 (initiated Jul 28, 2015)
StatusTerminated
ReasonBeckman Coulter, Inc. is recalling Access Thyroglobulin reagent packs because some may exhibit greater than 10% imprecision at concentrations <10 ng/mL.
Root causeUnder Investigation by firm
ActionOn 07/28/2015 the firm sent out customer notification letters via first class mail and e-mail. The firm's letter contained the following instructions: 1. Discard all Access Thyroglobulin reagent pack lots listed in the letter. 2. Evaluate Access Thyroglobulin results <10 ng/mL when using the reagent pack lot numbers listed to determine if controls and/or patients were higher than expected or did not match the clinical status of the patient. 3. At the discretion of the Laboratory Director, notify clinicians that it is possible the Access Thyroglobulin sample results reported by the laborator
Quantity3,035 (U.S.) and 7,123 (International)
DistributionWorldwide Distribution-US (nationwide) including Puerto Rico and the countries of Algeria, Australia, Austria, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Kore
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002905ACCESS THYROGLOBULIN REAGENTS ON THE ACCESS ANALYZER
Beckman Coulter, Inc., Oct 19, 2000
K031269ACCESS THYROGLOBULIN REAGENTS ON THE ACCESS IMMUNOASSAY SYSTEMS
Beckman Coulter, Inc., May 2, 2003