Recall Z-2488-2015 ยท MDI Prosthesis, Rib Replacement

Synthes (USA) Products LLC

VEPTR Vertical Expandable Prosthetic Titanium Rib, inferior cradles also known as caudal rib supports. The VEPTR/VEPTR II devices mechanically stabilize and/or correct thoracic deformities to allow the chest and lungs to grow.

DateAug 31, 2015 (initiated Jul 29, 2015)
StatusTerminated
ReasonThe last hole on certain VEPTR Inferior Cradles may be partially formed. Fully lengthening the construct to the partially formed last hole may result in mechanical failure of the construct due to rib sleeve breakage, and may require surgury to exchange components. Effects may also include pain, loss of deformity correction, and/or soft tissue/visceral injury.
Root causeProcess control
ActionUrgent recall letters, dated July 29, 2015, were sent to end users alerting them to the problem, the associated risk to patients, and provided instruction to return affected devices withthe response form.
Quantity3262
DistributionWorldwide distribution including US nationwide, Canada and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142587Vertical Expandable Prosthetic Titanium Rib
Depuy Synthes Spine, Inc., Nov 18, 2014