Recall Z-2479-2015 ยท KHP Acid, Lactic, Enzymatic Method

Siemens Healthcare Diagnostics Inc

RAPIDPoint 500 Blood Gas Analyzer, Siemens Material Numbers (SMN):10492730, 10696855, 10696857, 10697306 This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples.

DateAug 28, 2015 (initiated Jul 28, 2015)
StatusTerminated
ReasonWhen both ports (Serial and Ethernet) are configured to transmit data, the data stream from one port could potentially affect the data stream from the other port causing the data management system to include either: duplicate data, missing data, or data from a different patient.
Root causeSoftware design
ActionSiemens Healthcare issued Urgent Field Safety Notice sent by e-mail to all affected Siemens Healthcare Diagnostics Regional offices on July 28, 2015, both in the United States and outside of the United States, for communication with affected customers. This notice informs customers of the issue with the RAPIDPoint 500 dual port LIS transmission feature. Users are informed to: If the system is currently set up with dual LIS transmission enabled, one of the ports (either Serial or Ethernet) needs to be disabled. Instruction provided with the notification. Customer to complete Field Effectiven
Quantity5241 units
DistributionWorldwide Distribution - US (nationwide) Distribution and to the countries of : Albania, Algeria, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Bulgaria,Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002738BAYER RAPIDPOINT 400 SYSTEM
Bayer Corp., Dec 19, 2000
K020616BAYER RAPIDPOINT 405 SYSTEM
Bayer Diagnostics Corp., Jul 17, 2002
K110277RAPIDPOINT 405 NEONATAL BILIRUBIN
Siemens Healthcare Diagnostics, Aug 15, 2011
K113216RAPIDPOINT 500 BLOOD GAS ANALYER
Siemens Healthcare Diagnostics, May 3, 2012