Recall Z-2453-2015 ยท MVU Reagents, Specific, Analyte

Becton, Dickinson and Company, BD Biosciences

BD CD56 PE Cy 7: NCAM16.2 ASR Analyte Specific Reagent, Catalog number 335809 Hematology: The CD 56 antigen promotes homophilic adhesion in neurons and may be involved in homophilic adhesion of NK cells.

DateAug 24, 2015 (initiated Jul 17, 2015)
StatusTerminated
ReasonThree lots of CD56 PE Cy 7, Catalog number 335809, contain a low amount of HLA-DR PE antibody which could result in an unexpected staining pattern.
Root causeMaterial/Component Contamination
ActionBD Bioscience sent an Urgent Product Recall Letter dated July 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customer Actions : 1. If you have CD56 PE-Cy7 integrated into a Laboratory Developed Test (LDT) or panel, evaluate the impact of a false positive in the PE channel the interpretation of on the results. 2. Discard all remaining stock of CD56 PE-Cy7, Catalog Number 335809, lot number 4300708, 4304864, and/or 5013643, and request replacement product. BD will replace product at no cost. 3. As a requ
Quantity362 units
DistributionUS Distribution and Internationally to Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.