ClearanceBrief
Recall Z-2456-2015 ยท HTO Reamer

Trilliant Surgical Ltd.

Concave Reamer Head, 16 mm; Concave Reamer Head, 18 mm; Concave Reamer Head, 20 mm; Concave Reamer Head 22 mm; Convex Reamer Head, 16 mm; Convex Reamer Head, 18 mm; Convex Reamer Head 20 mm; Convex Reamer Head 22 mm, used with Gridlock Plating system, for use in trauma and reconstructive procedures in small bones.

DateAug 24, 2015 (initiated Jul 20, 2015)
StatusTerminated
ReasonThe cannulation of the Concave and Convex Reamer Heads may be too small to allow the Guide Pin to freely pass through.
Root causeComponent design/selection
ActionThe recalling firm sent out written notification of the recall on 7/3/14. The letter instructed the consignees to exchange affected instruments.
Quantity638 units
DistributionNationwide throughout the US
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121452GRIDLOCK PLATING SYSTEM
Trilliant Surgical, Ltd., Aug 20, 2012