Recall Z-2463-2015 ยท JSO Culture Media, Antimicrobial Susceptibility Test, Excluding Mueller Hinton Agar

Hardy Diagnostics

HardyCHROM MRSA Cat no: G307 Lot no: H15155 Expires: 2015-08-13 Container Type: 15x100mm monoplate Packaged: 10 plates/sleeve Storage: 2-8C on receipt. Light Sensitive. In vitro diagnostic www.HardyDiagnostics.com

DateAug 26, 2015 (initiated Jul 9, 2015)
StatusTerminated
ReasonHardy Diagnostics is recalling HardyCHROm MRSA due to a report of the potential for breakthrough of the methicillin-sensitive Staphylococcus aureus (MSSA) QC strain, ATCC 29213 on the identified lot.
Root causeUnknown/Undetermined by firm
ActionThe firm initially notified customers via phone to inform them of a recall. The phone scripts asks customers if they have had any issued with the QC testing. The firm states that our of a precaution they are asking customers to discard the remaining plates of the identified lot, and that credit or replacement plates will be offered for any remaining. The firm followed-up phone calls with a written notification letter. The letter states to complete, sign, and fax or email back the enclosed form stating compliance with the above action. The letters states that is the customer could not be r
Quantity2,820 (282 pk/10)
DistributionDistributed in the states of VA, PA, NY, NJ, IL, OH, and MD.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102922HARDYCHROM MRSA
Hardy Diagnostics, May 4, 2011