Recall Z-2339-2015 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare

GE Healthcare, Signa Infinity MR System; Signa MR/i MR System; SmartSpeed; EchoSpeed Plus; HighSpeed Plus, Signa Infinity MR System with EXCITE Technology, Excite II, Signa 3T Infinity with Excite, Signa Excite 1.5T and Signa Excite 3T, Signa Excite HD, Signa HD 1.5T and Signa HD 3.0T, Signa HDx, Signa HDxt, Signa HDi. Whole body magnetic resonance diagnostic imaging scanners.

DateAug 5, 2015 (initiated Jul 10, 2015)
StatusTerminated
ReasonA gradient cable in the ACGD cabinet may be missing a spacer that prevents connecting the gradient cable terminals in the wrong polarity on X gradient amplifier. This could cause images to be flipped left-right with incorrect orientation annotation.
Root causeNonconforming Material/Component
ActionGE Healthcare sent an Urgent Medical Device Correction letter dated July 10, 2015, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. Consignees with questions were instructed to contact GE Healthcare Service at 1-800-437-1171 or their local Service Representative.
Quantity82 (39 USA; 43 OUS)
DistributionWorldwide Distribution - US (nationwide) and Internationally to UNITED ARAB EMIRATES, ARGENTINA, BRAZIL, INDIA, EGYPT, CHINA, FINLAND, FRANCE, UNITED KINGDOM, HUNGARY, JAPAN, INDONESIA, KOREA, MEXICO, NEW ZEALAND, CANADA, AUSTRALIA, ITALY, RUSSIAN FEDERATION, TAIWAN, and SOUTH AFRICA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K0122001.5T INFINITY TWINSPEED MR SYSTEM
GE Medical Systems, Aug 3, 2001
K013636SIGNA INFINITY MR SYSTEM WITH EXCITE TECHNOLOGY
GE Medical Systems, Jan 24, 2002
K030874GE 3.0T SIGNA INFINITY TWINSPEED WITH EXCITE MR SYSTEM
General Electric Co., Mar 28, 2003
K040444GE SIGNA 3.0T WITH EXCITE MR SYSTEM
General Electric Co., Apr 14, 2004
K041476GE SIGNA EXCITE 1.5T MR SYSTEM, GE SIGNA EXCITE 3.0T MR SYSTEM
Ge Medical Systems, LLC, Jun 17, 2004