Recall Z-0005-2016 ยท KPS System, Tomography, Computed, Emission

Toshiba American Medical Systems Inc

Celesteion PCA-9000A/2 PET/CT System

DateOct 7, 2015 (initiated May 8, 2015)
StatusTerminated
ReasonIt was found that if specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved due to a software problem.
Root causeDevice Design
ActionPlanned Action by Toshiba Medical America Systems to address the defect: a. A notification will be sent to the customer with affected system in compliance with 21 CFR1 003.21 upon approval by the Secretary. b. To correct this issue revised software will be installed on the system to prevent this occurrence from happening. The consignee will be contacted by their Toshiba service representative for an appointment when the revised software becomes available. c. The customer will be advised to do the following until the corrective measures have been taken, i.e., revised software installed on thei
Quantity1
DistributionUS Distribution to the state of Nevada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140651CELESTEION, PCA-9000A/2
Toshibamedical Systems Corporation, Jun 11, 2014