Recall Z-2582-2015 ยท LRO General Surgery Tray

Customed, Inc

PLASTIC SURGERY PACK 5/CS Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. .

DateSep 3, 2015 (initiated Jul 10, 2015)
StatusTerminated
ReasonRecall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
Root causePackaging
ActionCustomed issued an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" notification letter dated July 10, 2015 to all affected customers. The letter identified the product, problem, and actions to be taken by the customer.
Quantity55 packs
DistributionDistributed Only in Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K904693CUSTOMED CARDIOVASCULAR PACK
Customed, Inc., Jan 2, 1991