Recall Z-2741-2015 ยท JFY Enzymatic Method, Creatinine

Siemens Healthcare Diagnostics, Inc.

Dimension CREA (Creatinine) and Dimension Vista CREA (Creatinine); The CREA method used on the Dimension clinical chemistry system is an in vitro diagnostic test intended for the quantitative determination of creatinine in human serum, plasma and urine.

DateSep 5, 2015 (initiated Jul 15, 2015)
StatusTerminated
ReasonSiemens has confirmed that the Dimension CREA and Dimension Vista CREA may produce falsely depressed results up to - 12.7% and -22% respectively, for a creatinine concentration of 1.5 mg/dL when hemolysis exceeds 200 mg/dL and 300 mg/dL respectively.
Root causeDevice Design
ActionSiemens sent an Urgent Medical Device Correction letter, dated July 2015, was provided to all Dimension CREA and Dimension Vista CREA customers who received the products to notify them of the issue, the potential risk to health, and actions to be taken by customers. If you have any questions please contact your Siemens Customer Care Center Technical Solutions at (800) 441-9250, or your local Siemens Technical Support Representative.
QuantityTotal number Dimension Vista - 32,165, Dimension - 238,634
DistributionWorldwide Distribution - US (nationwide) and to the countries of : Afghanistan, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia Herzegovinia, Bulgaria, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech R
FDA recordOpen on fda.gov

510(k) clearances named in this recall