Recall Z-2264-2015 ยท NPT Aortic Valve, Prosthesis, Percutaneously Delivered

Medtronic Cardiovascular Surgery-the Heart Valve Division

EnVeo R Loading System, Model No. LS-EnVeoR-23, LS-EnVeoR-2629, LS-EnVeoR-23-C, LS-EnVeoR-2629-C. Cardiovascular valve delivery system for use with the Medtronic CoreValve Evolut R system.

DateAug 5, 2015 (initiated Jul 15, 2015)
StatusTerminated
ReasonMedtronic is recalling the EnVeo R Loading Systems because of the presence of particulate.
Root causeProcess control
ActionAn Urgent Medical Device Recall Letter dated July 2015 will be be sent to customers who purchased the EnVeo R Loading System because Medtronic has received 8 reports related to this issue out of 7347 potentially affected units. The letter informs customers of the problems identified and the actions to be taken. Customers are instructed to complete the customer confirmation certificate and to email RS.CFQFCA@Medtronic.com or fax to Medtronic at (651) 367-0612 to the attention of Customer Focused Quality.
Quantity6,912 units total (540 units in US)
DistributionWorldwide Distribution - US Nationwide, Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Dominican Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Kuwait, Latvia, Lithuania, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Qatar, S