Recall Z-2433-2015 ยท LWK Razor, Surgical

Medline Industries Inc

Medline Electrical Clipper with Charging base The Electrical Surgical Clipper is intended for the removal of body hair from patients as required prior to surgery or other minor procedures.

DateAug 20, 2015 (initiated Jul 8, 2015)
StatusTerminated
ReasonCharging base of surgical clippers overheats and smokes.
Root causeProcess control
ActionMedline Industries sent a " Urgent Remedial Action Destruction Form " letter via first class dated July 9, 2015, to all affected customers.The letter identified the product the problem and the action needed to be taken by the customer. ***URGENT RECALL IMMEDIATE ACTION REQUIRED SURGICAL CLIPPER CHARGER BASE DYND70800 *** Medline Industries, Inc. is recalling specific lots of the Medline Surgical Clippers Charger Base (base only). This recall is being conducted due to a potential defective component on the circuit board causing the charger base to overheat which could result in a charger base
Quantity7062 units (U.S.); 540 units Foreign
DistributionWorldwide Distribution - US (nationwide) Distribution including Puerto Rico and to the countries of : United Arab Emirates, Barbados, Bahamas, Chile, Costa Rica, Italy, Lebanon, and Singapore.