ClearanceBrief
Recall Z-2193-2015 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems (Cleveland) Inc

GEMINI GXL 16 Slice CT/PET System. The Philips GEMINI PET/CT Imaging Systems are a family of integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) systems suitable for a wide range of diagnostic applications.

DateJul 22, 2015 (initiated Jul 1, 2015)
StatusTerminated
ReasonPhilips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.
Root causeSoftware design
ActionOn 7/6/2015 the firm sent Field Safety Notices to their customers.
Quantity2 units
DistributionWorldwide Distribution - US Nationwide, Australia, Belgium, Brazil, Canada, China, Denmark, France, Germany, Israel, Italy, Japan, Lebanon, Netherlands, Portugal, Saudi Arabia, South Africa, Switzerland, Thailand, Venezuela & Vietnam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051170GEMINI GXL
Philips Medical Systems (Cleveland), Inc., May 23, 2005