Recall Z-2200-2015 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips Electronics North America Corporation

Pinnacle3 Software Version 9.0, 9.2, 9.4 and 9.6, Model Numbers 453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451. UPDATED: Pinnacle3 Software Version 8.0h, 8.0k, 8.0m, 8.0n.

DateJul 22, 2015 (initiated Jul 31, 2014)
StatusTerminated
ReasonPhilips, Pinnacle Radiation Treatment Planning System version 8.0h, 8.0k, 8.0m, 8.0n, 9 0, 9 2 9 4, 9 6 is being recalled because the dose may be inconsistent with the density of a density-overridden ROl.
Root causeSoftware design
ActionConsignees were sent on 7/31/2014 a Philips "Urgent Field Safety Notice" dated 2014-July 20. The letter described the problem, product involved in the recall and the action that should be taken by the customers. For questions contact Customer Care Solutions Center at 1-800-722-9377. UPDATE: A second Philips letter dated 2 July 2015 was sent in August 6, 2015. This letter updated the recall to incorporate the additional products being recalled.
Quantity1642
DistributionWorldwide Distribution-US: All states in the US including DC and PR. OUS: Australia, Austria, Bangladesh, Belgium, Brazil, China, Denmark, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Malaysia, Mexico, Netherlands,New Zealand, Oman, Philippines, Singapore, South Africa, South
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041577PINNACLE3 RADIATION THERAPY PLANNING SYSTEM
Adac Laboratories, Jun 25, 2004